The flowmeter can be connected to a humidifier microbulle type with a 12×125 or 9/16″ thread.
Several combination are possible with flow (1 l/min to 15 l/min), fluids (air or oxygen) and the tank.
nanotube with humidifier microbulle
Design | Body and connection: nickel-plated brass Tank: glass, PC or PS |
|
Fluids | O2 or Air with sterile water for humidifier | |
Connection | 12×100 for probes according to NF, DIN, BS et UNI | |
Inlet pressure | 450 kPa ± 50 kPa | |
Outlet | Nipple 6 mm | |
Accuracy | ±10% or +/-0.5 l/min (values according to ISO15002) | |
Dimension/Weight | 315x64x90mm (hxlxw)/ 520g (250 ml glass) |